Veterinary Orthopedic Mini Drill Reaming Slow Speed Drill High Speed Drilling for Vet Small Bone

Product Description

E100 Electric Surgical Power Tools Orthopedic Multifunction Drill and Saw System
Product Description
Item Value
Multi-Function Handpiece Equipped with brushless DC canulated motor imported from USA
Battery Ni-MH with longer service-life, High capacity: 2100MAH, Voltage: 9.6V
Battery Box Peek material: sterilized more than 3000 times
Charger More convenient design to guarantee the battery life
Transfer Shield Imported stainless steel 304
Key Chuck Speed Range 0-1200rpm, Torque 3.5NM
Reaming Attachment Speed Range 0-400rpm, Torque 10NM
K-wire Attachment 0.7-1.6mm / 1.6-4.0mm
Reciprocating Saw Frequency: 0-12000cpm
Sterilizing Box 445*275*215mm, 4.6Kg
Veterinary Orthopedic Mini Drill Reaming Slow Speed Drill High Speed Drilling for Vet Small Bone
Veterinary Orthopedic Mini Drill Reaming Slow Speed Drill High Speed Drilling for Vet Small Bone
Veterinary Orthopedic Mini Drill Reaming Slow Speed Drill High Speed Drilling for Vet Small Bone
Veterinary Orthopedic Mini Drill Reaming Slow Speed Drill High Speed Drilling for Vet Small Bone
Key Features

High performance imported brushless motor

No-limit speed changing

Big capacity 2100mAh safe battery

High sealing system

One-hand operation

Perfect ergonomic design

Adapter integrated in the charger

Long service life. 2 years warranty for whole machine

Autoclavable at 135 °C (Improved type ≤ 155 °C), with 1.6Mpa

Standard Configuration
1.
One handpiece
2.
One battery
3.
One charger
4.
One key chuck or reaming attachment
5.
One sterilization channel
6.
One aluminum alloy case
Warranty
Whole machine and motor: 2 years.
Battery and fast-consumable items: 3 months.
Operation & Maintenance Notes
Important Safety Instructions
1.
Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C). If the temperature exceeds 135 °C, the motor will be damaged.
2.
Normal bone drill can only hold drill bits less than 6mm. Acetabulum reaming drill works with big torque at around 300 rpm. Canulate drill is used with nail or K-wire at around 600 rpm.
3.
The battery needs to be removed from the power tool after surgery. Otherwise, the automatic micro-discharge will shorten the battery service life.
4.
The battery cannot be sterilized and is not autoclavable.
5.
The cranial bur must be washed immediately after surgery by purger to avoid blood sticking inside, then kept dry.
6.
Ensure there is no water remaining in the battery space of the handpiece after sterilization.
Related Products
Veterinary Orthopedic Mini Drill Reaming Slow Speed Drill High Speed Drilling for Vet Small Bone
Certifications
Certification 1
Certification 2
Certification 3
Certification 4
Packaging & Shipping
Packaging & Shipping details
FAQ Info Chart
FAQ Details Table
Frequently Asked Questions

What is the warranty coverage for this surgical drill system?

The warranty for the whole machine and motor is 2 years. For the battery and other fast-consumable parts, the warranty is 3 months.

Can the battery pack be sterilized in an autoclave?

No, the battery pack cannot be sterilized and is not autoclavable. It must be handled with appropriate sterile transfer shields during surgical procedures.

What is the maximum autoclavable temperature for the handpiece?

The standard sterilizing temperature should not exceed 135 °C. For the improved version, it can handle up to 155 °C. Exceeding these temperatures will cause damage to the motor.

How should the battery be maintained after surgery?

The battery must be removed from the power tool handpiece immediately after the operation. Leaving it inside will cause automatic discharge, reducing the overall lifespan of the battery.

What are the speed ranges for the attachments?

The key chuck attachment runs at a speed range of 0-1200rpm (torque 3.5NM), while the reaming attachment operates at 0-400rpm (torque 10NM) to provide high-torque power.

What components are included in the standard configuration package?

The standard kit includes one handpiece, one battery, one charger, one key chuck or reaming attachment, one sterilization channel, and one protective aluminum alloy case.

Related Products